Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Magnetom Symphony syngo, Model Number: 7104594, Siemens Medical Solutions USA, Malvern PA, 19355

Recall number
Z-1047-2008
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Magnetom Symphony syngo, Model Number: 7104594, Siemens Medical Solutions USA, Malvern PA, 19355
Status
Terminated
Initiated
2007-12-01
Posted
2008-03-25
Terminated
2012-10-01
Reason
Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
Root cause
Device Design
Action
The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
Quantity
141 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Serial Numbers: 22021, 22024, 22060, 22074, 22079, 22108, 22115, 22116, 22121, 22127, 22172, 22196, 22225, 22242, 22272, 22306, 22311, 22342, 22352, 22353, 22356, 22402, 22440, 22441, 22446, 22471, 22498, 22510, 22520, 22523, 22530, 22561, 22565, 22596, 22598, 22604, 22619, 22621 22622, 22636, 22674, 22690, 22698, 22711, 22801, 22810, 22825, 22835, 22836, 22865, 22881, 22885, 22907, 22911, 22914, 22916, 22944, 22967, 23001, 23057, 23059, 23085, 23099, 23104, 23132, 23136, 23156, 23176, 23199, 23218, 23247, 23252, 23255, 23284, 23315, 23323, 23357, 23381, 23491, 23497, 23507, 23538, 23552, 23574, 23578, 23579, 23580, 23585, 23590, 23594, 23596, 23602, 23604, 23605, 23610, 23613, 23628, 23641, 23642, 23672, 23682, 23687, 23691, 23698, 23731, 23760, 23762, 23764, 23782, 23783, 23788, 23796, 23804, 23823, 23834, 23839, 23883, 23885, 23902, 23904, 23913, 23919, 23932, 23943, 23950, 23956, 23984, 23986, 23987, 23994, 23997, 23999, 24017, 24019, 24022, 24030, 24043, 24071, 24099, 24104, and 24107.
510(k) numbers
["K970852"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register