Abbott Laboratories HPD/ADD: AxSYM AFP Reagent Pack, list 7A48-22 (100 test kit), 7A48-21(French label); and AxSYM Cancer Assay Disk, list 3D50-02 (Version 2.0), 3D50-03 (Version 3.0), and 3D50-04 (Version 4.0)
- Recall number
- Z-1047-03
- Recalling firm
- Abbott Laboratories HPD/ADD
- Product
- AxSYM AFP Reagent Pack, list 7A48-22 (100 test kit), 7A48-21(French label); and AxSYM Cancer Assay Disk, list 3D50-02 (Version 2.0), 3D50-03 (Version 3.0), and 3D50-04 (Version 4.0)
- Status
- Terminated
- Initiated
- 2003-06-12
- Posted
- 2003-07-24
- Terminated
- 2003-11-21
- Reason
- AFP results obtained when using the 1:101 automated dilution protocol showed an overestimation of AFP of up to 30.5%
- Root cause
- Other
- Action
- All AxSYM AFP customers were informed not to use the 1:101 automated dilution protocol via Device Correction letter dated 6/12/03. If necessary, manually dilute samples per the AxSYM AFP package insert.
- Quantity
- 23,000 kits
- Distribution
- Nationwide and internationally to Australia, Canada, Costa Rica, Ecuador, Japan, Panama, San Salvador, Thailand, Barbados, Chili, Curacao, Honduras, Korea, Peru, Singapore, United Kingdom, Brazil, Columbia, Dominican Republic, Hong Kong, Pakistan, St. Vincent, Uruguay
- Lot, serial or model codes
- All lots of AFP list 7A48-22 and 7A48-21 used with all lots of AxSYM Cancer Assay Disk list 3D50-02, 3D50-03 and 3D50-04 with AFP Assay File version 2.00.200, when used with AxSYM SYstem Sotware Versions 3.00 and higher.
- 510(k) numbers
- not on file
- PMA numbers
- ["P820060S011"]
- Product code
- LOK
- Firm FEI
- 1415939
- Firm city
- Abbott Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28