Skip to the content

FDA Medical Device Recalls (openFDA)

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: SMV BodyTrack, System, Tomography, Computed, Emission

Recall number
Z-1046-2026
Recalling firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Product
SMV BodyTrack, System, Tomography, Computed, Emission
Status
Open, Classified
Initiated
2025-12-10
Posted
2026-01-09
Terminated
not on file
Reason
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Root cause
Package design/selection
Action
GE Healthcare began notifying their consignees by telephone on 12/10/2025, followed by an URGENT MEDICAL DEVICE CORRECTION UPDATE notice distributed on 01/05/2025 via US mail. The notice explained the issue, safety risk, and requested the consignee stop use of the product. Care should be exercised when deinstalling these systems. To discuss replacement options, your GE HealthCare account representative. Should be contacted.
Quantity
12 units
Distribution
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
Lot, serial or model codes
UDI/DI Not applicable: All serial numbers in distribution
510(k) numbers
["K911346"]
PMA numbers
not on file
Product code
IYX
Firm FEI
3004571294
Firm city
Tirat Carmel
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register