Skip to the content

FDA Medical Device Recalls (openFDA)

Arjo, Inc.: Alenti Lift Hygiene Chair; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model numbers CDB8003-01-US and CDB8053-01-US (with scale)

Recall number
Z-1046-04
Recalling firm
Arjo, Inc.
Product
Alenti Lift Hygiene Chair; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model numbers CDB8003-01-US and CDB8053-01-US (with scale)
Status
Terminated
Initiated
2004-06-08
Posted
2004-07-20
Terminated
2004-12-13
Reason
There has been an increased number of tipping incidents and subsequent patient injuries.
Root cause
Other
Action
Safety Advice Notice letters dated 6/8/04 were sent by certified mail to the end users, informing them of the possibility of tipping if the Alenti catches on other equipment; if the resident is not positioned correctly on the chair; if the Alenti is used with an inappropriate resident; if the resident is left unattended, particularly when the Alenti is in an elevated position; and if the safety belt is not used as a positioning belt. The end users were provided with a laminated hang tag to place on the lift with the listed safety aspects and warnings, a copy of the revised Alenti Operating and Product Care Instructions dated April 2004, and an Alenti safety belt for each of their lifts. The users were requested to complete enclosed customer response form indicating that they received and understood the safety advice notice, that they received the revised manual and destroyed all previous copies of the manual, that they received a safety belt for each lift, and that they made the patient lift operators aware of the safety advice notice and that they understood its contents. Any questions were directed to the Quality Assurance Department, at 1-800-323-1245, ext. 6103. Revised hang tags including instructions to check the castors on the lift to assure that they are secured and functioning properly, as well as providing the operating instructions for the safety belts, were sent to the accounts on 7/13/04.
Quantity
1,035 units
Distribution
Nationwide and Canada
Lot, serial or model codes
model numbers CDB8003-01-US and CDB8053-01-US, all serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
1419652
Firm city
Roselle
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register