Skip to the content

FDA Medical Device Recalls (openFDA)

The Binding Site Group, Ltd.: Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

Recall number
Z-1043-2017
Recalling firm
The Binding Site Group, Ltd.
Product
Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.
Status
Terminated
Initiated
2016-07-05
Posted
2017-01-18
Terminated
2017-07-03
Reason
Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.
Root cause
Under Investigation by firm
Action
On 07/05/16 notification letters were sent to customers. The Binding Site Group requests the following: - Immediately Quarantine & Dispose of any LK018.OPT lot 405566 kits within stock; - Remove any affected lots from the Optilite instrument; - Patients with Lambda Freelite results >114mg/L should be reviewed in consultation with clinicians, alongside other clinical findings and retested where applicable; - Notify local Binding Site representative to request replacement kits; - Return completed E-Back Form to local Binding Site Representative; The firm recommended that the notice be circulated to those within the organization who need to be aware.
Quantity
809 kits
Distribution
U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.
Lot, serial or model codes
Lot #405566, 405567
510(k) numbers
["K150658"]
PMA numbers
not on file
Product code
DEH
Firm FEI
3002808340
Firm city
Birmingham
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register