Smith & Nephew, Inc.: JOURNEY II XR TIBIAL POSTERIOR KEEL PUNCH SZ 1-2 , REF 74013986 The JOURNEY II XR Tibial Posterior Keel Punch is a reusable surgical instrument used to prepare the proximal tibial to receive a JOURNEY II XR tibial baseplate.

Recall number
Z-1042-2019
Recalling firm
Smith & Nephew, Inc.
Product
JOURNEY II XR TIBIAL POSTERIOR KEEL PUNCH SZ 1-2 , REF 74013986 The JOURNEY II XR Tibial Posterior Keel Punch is a reusable surgical instrument used to prepare the proximal tibial to receive a JOURNEY II XR tibial baseplate.
Status
Terminated
Initiated
2018-12-10
Posted
not on file
Terminated
2020-11-03
Reason
Higher than anticipated occurrence of bone fracture during the use of the XR Tibia Posterior Punch
Root cause
Device Design
Action
The firm, smith&nephew, initiated the recall by email with an "Urgent Medical Device Recall Notice" letter dated 12/10/18 to the customers on 12/10/2018. Letter describes the product, problem and actions to be taken. The customers were instructed to do the following: -Inspect your inventory and locate any devices from the listed product and batch numbers and quarantine them immediately. - If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. -Complete and return the Response Form to FieldActions@smith-nephew.com or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
Quantity
65 units
Distribution
US Distribution to OR and TN; and Internationally to: Japan.
Lot, serial or model codes
Batch Numbers: 17JGA0020; 17JGA0020A; 17JGA0020B; 17JGA0026; 17JGA0026A; 17JGA0026B; 17JGA0032A; 17JGA0032B; 17JGA0042; 17JGA0042A; 17JGA0042R
510(k) numbers
["K173331"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register