Skip to the content

FDA Medical Device Recalls (openFDA)

Lang Dental Mfg Co, Inc.: Ortho-Jet Powder, fast curing orthodontic acrylic resin; the product was packaged under the following brands and sizes: a.1) Ortho-Jet Powder, Lang Dental Manufacturing Co., Inc., Wheeling, IL 60090-0969, product code 1320, 100 g plastic squeeze bottle; a.2) Biocryl Resin Acrylic, Great Lakes Orthodontics, Ltd., P.O. Box 5111, Towanda, New York 14151-5111, product code 040-016, 1 lb. lab pack with empty plastic squeeze bottle; b.1) Contemporary Colored Ortho-Jet Powder, Lang Dental, product code 1520, 100 g plastic squeeze bottle; b.2) Contemporary Colored Assorted Kit, Great Lakes Orthodontics, product code 045-088, 2 oz. plastic squeeze bottles, 6 bottles per kit; c.1) Electric Colored Ortho-Jet Powder, Lang Dental, product code 2720, 100 g plastic squeeze botles and product code 2793, 45 g plastic squeeze bottles/6 per kit; c.2) Tropical Tones Acrylic Resin, Great Lakes Orthodontics, product code 045-052, 4 oz. plastic squeeze bottle

Recall number
Z-1042-04
Recalling firm
Lang Dental Mfg Co, Inc.
Product
Ortho-Jet Powder, fast curing orthodontic acrylic resin; the product was packaged under the following brands and sizes: a.1) Ortho-Jet Powder, Lang Dental Manufacturing Co., Inc., Wheeling, IL 60090-0969, product code 1320, 100 g plastic squeeze bottle; a.2) Biocryl Resin Acrylic, Great Lakes Orthodontics, Ltd., P.O. Box 5111, Towanda, New York 14151-5111, product code 040-016, 1 lb. lab pack with empty plastic squeeze bottle; b.1) Contemporary Colored Ortho-Jet Powder, Lang Dental, product code 1520, 100 g plastic squeeze bottle; b.2) Contemporary Colored Assorted Kit, Great Lakes Orthodontics, product code 045-088, 2 oz. plastic squeeze bottles, 6 bottles per kit; c.1) Electric Colored Ortho-Jet Powder, Lang Dental, product code 2720, 100 g plastic squeeze botles and product code 2793, 45 g plastic squeeze bottles/6 per kit; c.2) Tropical Tones Acrylic Resin, Great Lakes Orthodontics, product code 045-052, 4 oz. plastic squeeze bottle
Status
Terminated
Initiated
2004-06-07
Posted
2004-07-20
Terminated
2004-09-20
Reason
The dental acrylic resin powders are contaminated with small black particles believed to be polyethylene.
Root cause
Other
Action
Lang Dental sent recall letters dated 6/7/04 to their direct accounts, informing them of the contamination of the acrylic resin powders with the black particles and requesting the return of any existing inventory of the affected lots of product for replacement. Distributors were requested to sub-recall the products from their accounts. Follow-up letters were sent to the accounts on 6/14/04, clarifying the lot numbers affected to include both the lot number on the box and the lot number on the bottom of the squeeze bottles affected.
Quantity
45 units
Distribution
Nationwide
Lot, serial or model codes
a.1) product code 1320, box lot 1320-4005, bottle lot 60274005/04AB;  a.2) product code 040-016, lot 53-1334-4016, empty bottle and cap only;  b.1) product code 1520, box lot 1520-4020, bottle lot 15144002/01AB - Blue; box lot 1520-4023, bottle lot 15244006/01AH - White;  b.2) product code 045-088, box lot 53-1593-4006, bottle lots 15224001/01AQ - red, 15203002/01BS - orange, 15144002/01AA - blue and 15124003/01AP - black, and box lot 53-1593-4007, bottle lots 15124003/01AQ - black, 15144002/01AC - blue, 15203002/01BU - orange, 15224001/01AT - red, 15244006/01AF - white, 15263001/01CD - yellow;  c.1) product code 2720, box lot lot 2720-4021, bottle lot 60374001/01AH -pink; box lot 2720-4022, bottle lot 60363002/01BE -yellow; box lot 2720-4023, bottle lot 60374001/01AK - pink;  product code 2793, kit box lot 2793-4005, bottle lots 60343002/02AF - green, 27124001/01AF - orange, 60374001/01AJ - pink, 27164001/01AS - red, 60363002/01BD - yellow and 60353005/01AS - blue;  c.2) product 045-052, box lot 53-2720-4044, bottle lot 60343002/02AE - key lime green
510(k) numbers
not on file
PMA numbers
not on file
Product code
EBG
Firm FEI
1417322
Firm city
Wheeling
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register