Skip to the content

FDA Medical Device Recalls (openFDA)

Mobius Imaging, LLC: Detachable 10in Pendant Upgrade component for the AIRO Computed Tomography (CT) X-ray System.

Recall number
Z-1041-2026
Recalling firm
Mobius Imaging, LLC
Product
Detachable 10in Pendant Upgrade component for the AIRO Computed Tomography (CT) X-ray System.
Status
Open, Classified
Initiated
2025-12-04
Posted
2026-01-09
Terminated
not on file
Reason
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
Root cause
Labeling Change Control
Action
An URGENT MEDICAL DEVICE CORRECTION notice dated 12/4/25 instructs consignees to ensure the following mitigations are in place while the correction is deployed: the hand position warnings as indicated in the IFU, train operators, the AIRO is accessible through a physical key switch and password protected log in screen, and the Backing Up System Warning label is affixed on to the gantry of affected devices. Stryker will correct devices during an upcoming service visit. Consignees are to inform users of devices of this recall by sharing the recall notice, immediately check stock areas and operating rooms to identify affected devices, and inform all users of the potential harms and their risk mitigations. Consignees with any questions are to contact their local sales office or Stryker Representative.
Quantity
26 units
Distribution
US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.
Lot, serial or model codes
Part No. MI-70-0128; UDI: (01) 00869346000200 (11)230920 (21) AIRO-XXXX; Serial No. 2232206183, 2232206193, 2232206203, 2232206213, 2232206223, 2232206233, 2235522143, 2235522153, 2235522163, 2236339153, 2236339163, 2236339173, 2236339183, 2303429483, 2303429503, 2313925663, 2313925673, 2316030773, 2316030783, 2316703503, 2316703523, 2317415473, 2320006923, 2320006933, 2323513413, 70-0128-13904392.
510(k) numbers
["K180393"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3010151377
Firm city
Shirley
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register