Skip to the content

FDA Medical Device Recalls (openFDA)

The Anspach Effort, Inc.: Anspach Xmax Pneumatic Motor System; ENT Surgical Drill, including handpiece and Accessories Attached. All Models; Black Motors; Anspach Effort, Inc., Palm Beach FL 33410.

Recall number
Z-1041-2008
Recalling firm
The Anspach Effort, Inc.
Product
Anspach Xmax Pneumatic Motor System; ENT Surgical Drill, including handpiece and Accessories Attached. All Models; Black Motors; Anspach Effort, Inc., Palm Beach FL 33410.
Status
Terminated
Initiated
2006-12-18
Posted
2008-02-13
Terminated
2012-07-06
Reason
Incorrect Sterilization Process --The Operating Manual for the product contains an incorrect statement regarding the sterilization of the Autolube foot control. The manual states that the product can withstand ETO sterilization, but there is no data to support this process.
Root cause
Labeling False and Misleading
Action
Anspach issued an Urgent Device Correction letter on 11/27/06, to all manual holders advising them of the error and requested that they update the manual with the corrected instructions. A reply receipt confirming notification was included.
Quantity
7,399 total units
Distribution
Worldwide Distribution
Lot, serial or model codes
All Serial Numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
ERL
Firm FEI
1045834
Firm city
Palm Beach Gardens
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register