Skip to the content

FDA Medical Device Recalls (openFDA)

Telefelx Medical: Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and 525-30.

Recall number
Z-1040-2010
Recalling firm
Telefelx Medical
Product
Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and 525-30.
Status
Terminated
Initiated
2010-01-12
Posted
2010-03-09
Terminated
2011-04-05
Reason
When turned on, the Neptune heater may cause an artifact on the patient's heart monitor.
Root cause
Device Design
Action
Teleflex consignees were notified via Urgent Medical Device Recall letter via next day delivery on/about 01/12/2010. They were asked to return all of the affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. A second mailing to non-responders on the consignee list will be made approximately four to six weeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or as an alternative, they may be contacted by phone, e-mail or fax. A record of these contacts will be maintained. Affected products received by Teleflex Medical will be reworked with software upgrade and placed back into distribution. Consignees will either receive a software upgrade on site by a third party or replacement product with the software upgrade. Records will be maintained accordingly.
Quantity
15,134 units
Distribution
U.S. Canada, India, Japan, Puerto Rico, Germany, France, Italy and Ireland.
Lot, serial or model codes
Product numbers: 425-00, 425-10 and 525-30, All lot and serial numbers.
510(k) numbers
["K063758"]
PMA numbers
not on file
Product code
BTT
Firm FEI
3005747797
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register