FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item Number 00-5994-022-14
- Recall number
- Z-1038-2025
- Recalling firm
- Zimmer, Inc.
- Product
- NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item Number 00-5994-022-14
- Status
- Open, Classified
- Initiated
- 2025-01-06
- Posted
- 2025-01-28
- Terminated
- not on file
- Reason
- It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.
- Root cause
- Process change control
- Action
- An URGENT MEDICAL DEVICE RECALL - REMOVAL notification letter dated 1/6/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to orporateQuality.PostMarket@zimmerbiomet.com. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products available to return in your territory. b. For International Returns, request an IRA by emailing zimmerbiometintlirareques
- Quantity
- 155 units
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN.
- Lot, serial or model codes
- UDI-DI: 00889024635647; Lot Numbers: 66602503 66520665 66881918 66949906
- 510(k) numbers
- ["K960279"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28