FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc.: Dimension Vista V-LYTE Fluids: Standard B/Salt Bridge catalog number K825
- Recall number
- Z-1038-2010
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Dimension Vista V-LYTE Fluids: Standard B/Salt Bridge catalog number K825
- Status
- Terminated
- Initiated
- 2009-10-22
- Posted
- 2010-03-10
- Terminated
- 2010-12-16
- Reason
- Results can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor.
- Root cause
- Other
- Action
- The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
- Quantity
- 150 units
- Distribution
- The product was shipped to medical facilities nationwide. The product was also shipped to Australia, Canada, Belgium, France, Germany, Italy, Japan, Malaysia, New Zealand, Netherlands, Portugal, Spain, South Korea, and Switzerland.
- Lot, serial or model codes
- Catalog number K825
- 510(k) numbers
- ["K051087"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28