Siemens Medical Solutions USA, Inc: Leonardo Workstation Picture Archiving and Communication System, Model Number: 7129534, Software Version: VB30B, Siemens Medical Solutions, Inc., Malvern, PA 19355
- Recall number
- Z-1038-2008
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Leonardo Workstation Picture Archiving and Communication System, Model Number: 7129534, Software Version: VB30B, Siemens Medical Solutions, Inc., Malvern, PA 19355
- Status
- Terminated
- Initiated
- 2007-12-21
- Posted
- 2008-04-29
- Terminated
- 2008-09-30
- Reason
- Incorrect display: The orientation labels will be incorrectly displayed on the reconstructed InSpace 3-D image if the orientation was not originally HFS on the acquisition system.
- Root cause
- Software design
- Action
- Siemens Medical Solutions issued a Customer Safety Advisory Notice dated 12/21/07 to its customers to inform them of the problem and instructions to avoid the problem until the firm's representative is available to upgrade the software.
- Quantity
- 47 Units
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Model Number: 7129534. Serial Numbers: 10483, 10651, 10628, 10488, 10605, 10676, 10658, 10421, 10685, 10724, 10584, 10619, 10626, 10629, 10635, 10694, 10706, 10438, 10600, 10668, 10655, 10436, 10606, 10702, 10433, 10558, 10636, 10642, 10692, 10698, 10412, 10440, 10539, 10613, 10674, 10621, 10634, 10583, 10684, 10476, 10414, 10664, 10671, 10712, 10472, 10567, and 10641. Software version VB30B.
- 510(k) numbers
- ["K040970"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28