Siemens Medical Solutions USA, Inc: Leonardo Workstation Picture Archiving and Communication System, Model Number: 7129534, Software Version: VB30B, Siemens Medical Solutions, Inc., Malvern, PA 19355

Recall number
Z-1038-2008
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Leonardo Workstation Picture Archiving and Communication System, Model Number: 7129534, Software Version: VB30B, Siemens Medical Solutions, Inc., Malvern, PA 19355
Status
Terminated
Initiated
2007-12-21
Posted
2008-04-29
Terminated
2008-09-30
Reason
Incorrect display: The orientation labels will be incorrectly displayed on the reconstructed InSpace 3-D image if the orientation was not originally HFS on the acquisition system.
Root cause
Software design
Action
Siemens Medical Solutions issued a Customer Safety Advisory Notice dated 12/21/07 to its customers to inform them of the problem and instructions to avoid the problem until the firm's representative is available to upgrade the software.
Quantity
47 Units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Model Number: 7129534. Serial Numbers: 10483, 10651, 10628, 10488, 10605, 10676, 10658, 10421, 10685, 10724, 10584, 10619, 10626, 10629, 10635, 10694, 10706, 10438, 10600, 10668, 10655, 10436, 10606, 10702, 10433, 10558, 10636, 10642, 10692, 10698, 10412, 10440, 10539, 10613, 10674, 10621, 10634, 10583, 10684, 10476, 10414, 10664, 10671, 10712, 10472, 10567, and 10641. Software version VB30B.
510(k) numbers
["K040970"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register