FDA Medical Device Recalls (openFDA)

ConvaTec, Inc: Nelaton Urinary Catheter, disposable, sterile, 100 eaches per sales unit. Labeled as: CH16/40CM and CH14/40CM Product Usage: Urinary Catheters are sterile disposable devices intended to provide an intermittent pathway for drainage of fluids from the bladder. They are intended to be inserted intermittently through the urethra and up to the bladder enabling urine to drain. They are designed for short-term, intermittent transient use only and are not intended to be left in situ (there is no balloon on the tip to allow the catheter to be retained within the bladder). They are intended for single use only. They are intended to be removed once the bladder is fully drained. The goal of intermittent catheterization using Urology Catheters is to completely empty the bladder and prevent Urinary Tract Infections (UTIs).

Recall number
Z-1037-2019
Recalling firm
ConvaTec, Inc
Product
Nelaton Urinary Catheter, disposable, sterile, 100 eaches per sales unit. Labeled as: CH16/40CM and CH14/40CM Product Usage: Urinary Catheters are sterile disposable devices intended to provide an intermittent pathway for drainage of fluids from the bladder. They are intended to be inserted intermittently through the urethra and up to the bladder enabling urine to drain. They are designed for short-term, intermittent transient use only and are not intended to be left in situ (there is no balloon on the tip to allow the catheter to be retained within the bladder). They are intended for single use only. They are intended to be removed once the bladder is fully drained. The goal of intermittent catheterization using Urology Catheters is to completely empty the bladder and prevent Urinary Tract Infections (UTIs).
Status
Open, Classified
Initiated
2019-01-09
Posted
not on file
Terminated
not on file
Reason
An internal assessment of packaging confirmed the potential for a breach in the sterile barrier.
Root cause
Package design/selection
Action
ConvaTec notified customers on about 01/09/2019 via "URGENT: VOLUNTARY MEDICAL DEVICE RECALL" letter. Instructions included how to identify affected product on the primary packaging and/or shipping carton as well as specific instructions for distributors, retailers, and end users (hospitals, EMT services, others). Distributors were instructed to inspect their inventory for affected products, perform a count of affected product in inventory, complete and return the response form, contact Customer Service at 1-800-582-6514 or ushcreturns@ups.com to report affected inventory on hand, forward the recall notice package to customers if product was further distributed, and provide a complete list of all consignees to allow ConvaTec to perform effectiveness checks. Retailers were instructed to immediately stop distributing and quarantine any affected inventory, perform a count of affected product in inventory, complete and return the response form, contact Customer Service at 1-800-582-6514 or ushcreturns@ups.com to report affected inventory on hand, and post page one of the recall notice in a conspicuous location in the store. End users (hospitals, EMT service, others) were instructed to inspect inventory to confirm if any affected product codes are in inventory, perform a count of affected product in inventory, complete and return the response form, contact Customer Service at 1-800-422-8811 or ushcreturns@ups.com to report affected inventory on hand.
Quantity
5,744,525 units total
Distribution
Worldwide Distribution - US Nationwide in the states of IL, MI, PA, WI. and the countries of Algeria, Australia, Austria, Belgium, Canada, Croatia, Czechia, Czech Republic, Denmark, Egypt, Estonia, Finland, Faroe Islands, France, Germany, Greece, Hong Kong, Hungary, Iceland, Israel, Italy, Kenya, Korea, Kuwait, Lebanon, Myanmar, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Spain, Slovakia, South Africa, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, United States, Vietnam.
Lot, serial or model codes
Product Codes/REF Numbers: 505599, 505603; Lot Codes: 222348, 222349
510(k) numbers
["K896729"]
PMA numbers
not on file
Product code
KOD
Firm FEI
3011987967
Firm city
Greensboro
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register