FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp. (NeuroSciences): Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.
- Recall number
- Z-1036-2026
- Recalling firm
- Integra LifeSciences Corp. (NeuroSciences)
- Product
- Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.
- Status
- Open, Classified
- Initiated
- 2025-12-12
- Posted
- 2026-01-08
- Terminated
- not on file
- Reason
- Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.
- Root cause
- Process control
- Action
- Integra LifeSciences Corporation notified consignees on about 12/12/2025 via FedEx overnight mailing. Consignees were instructed to identify impacted part numbers and lot numbers on hand, remove them immediately from service and quarantine them, complete and return the Acknowledgement Form, and maintain the notice for records. Consignees were also instructed to forward the notification to those who utilize the product for awareness of the recall. Once the Acknowledgement Form is received, and noted that affected product are on hand, Customer Service will contact consignees and provide a Return Material Authorization (RMA) number and directions to return the affected product.
- Quantity
- 1,301 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
- Lot, serial or model codes
- Catalog Number: 626631US. UDI-DI: 10381780514473. Lot Numbers: 7288007, 7288040, 7294017, 7305988, 7305989, 7305990, 7305991, 7307683, 7325489, 7336088, 7336089, 7336090, 7305992, 7346223, 7391752, 7391751, 7406650, 7406651, 7406652, 7408825, 7408826, 7408827, 7456174, 7456175, 7456176, 7456179, 7478039, 7478040, 7480073, 7480072, 7482162, 7503704, 7507495, 7503608, 7510153, 7510155, 7510154, 7534856, 7535008, 7535009, 7411809.
- 510(k) numbers
- ["K153347"]
- PMA numbers
- not on file
- Product code
- GWM
- Firm FEI
- 3003418325
- Firm city
- Princeton
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as