FDA Medical Device Recalls (openFDA)
Olympus Scientific Solutions Americas: Olympus Scientific Solutions Americas Corporation (OSSA) VANTA XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems
- Recall number
- Z-1036-2017
- Recalling firm
- Olympus Scientific Solutions Americas
- Product
- Olympus Scientific Solutions Americas Corporation (OSSA) VANTA XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems
- Status
- Terminated
- Initiated
- 2017-01-07
- Posted
- not on file
- Terminated
- 2019-08-27
- Reason
- It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.
- Root cause
- Radiation Control for Health and Safety Act
- Action
- Olympus will bring the defect compliance : 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office recall coordinator noted below are to be included in the notification. 2. Tracking of customer/owner self-installation(s) of the software update should be done to the maximum possible extent via serial-number-tracked downloads, follow-up email reporting, return-mail card or by other similarly effective means. For further questions, please call (781) 419-3500.
- Quantity
- US 74
- Distribution
- US Distribution to the states of : PA, MA, CO, OH, NJ, GA, CA, TX, IL, VA, MN, NH, WA, SC, AZ, LA, MD, OK, IN, NY, NC and KS.
- Lot, serial or model codes
- affects all units.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- RBY
- Firm FEI
- 3006241029
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as