GE Medical Systems, LLC: GE Innova 4100/4100 IQ. , GE Healthcare, Waukesha, WI 53188 ( The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector). The product is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures and optionally, rotational imaging procedures. It is also intended for generating fluoroscopic images of human anatomy for cardiology, diagnostic and interventional procedures.

Recall number
Z-1036-2009
Recalling firm
GE Medical Systems, LLC
Product
GE Innova 4100/4100 IQ. , GE Healthcare, Waukesha, WI 53188 ( The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector). The product is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures and optionally, rotational imaging procedures. It is also intended for generating fluoroscopic images of human anatomy for cardiology, diagnostic and interventional procedures.
Status
Terminated
Initiated
2008-11-04
Posted
2009-09-24
Terminated
2012-05-16
Reason
One of three copper spectral filters could remain within the x-ray while the system displays no spectral filter. The displayed Air Kerma rate and thus cumulative Air Kerma are overestimated. The Air Kerma rate may be over estimated by up to 60%.
Root cause
Radiation Control for Health and Safety Act
Action
Affected customers were notified by means of a Important Electric Product Radiation Warning letter. GE field personnel will be replacing the collimators.
Quantity
166
Distribution
Nationwide Distribution -- AL, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and Puerto Rico.
Lot, serial or model codes
system ID: 201541SP1, 207662OR4100, 209467INNOVA308, 212639ANI4100, 215537INNOVA1, 215662INNOVA1, 216778ANGIO1, 216778ANGIO2, 2198864100, 228809VASC, 229312IN41OR, 229434INNOV, 239939INNOVA, 248849SP1, 248937IN1, 252847PRS2, 252847PRS3, 256429SV1, 281401T4100, 303399IR, 303436ANGIO, 303436ANGIO2, 304256INNOVA, 304347INNOVA2, 304424CCINOVA, 3046234100, 305823IN4100, 305835IN41, 309343CSP1A, 309655FSP1A, 309655FSP3A, 309672MCL2, 313343SP1, 314966DPHSP, 316651INNOVA41, 316962INNOVA41, 317621PV, 317887INNOVAPV, 318681INNOVA1, 330489ORVASC, 330489SPEC9, 386254HS1, 386254HS2, 386HPOS1, 404466SF41A, 404712SP12, 404712SP13, 404GHSIR1, 404GHSIR2, 409212INOV41, 410337INNOVA2, 412641IR, 412647OR10, 412784LAB1, 412784LAB2, 4134474100, 4135344100, 414649CL2, 414649DVI, 419557ORLAB, 423495ENDO2, 423495MSPEC2, 443481ANGIO1, 4436434100, 4438434100, 478765ANG1, 480512SP, 5033704100, 505848CATH2, 505848V4100, 508650IN41, 509474SH41, 509482HFI, 509575YM4100, 512528CPCL1, 512901NI4100, 5135844100, 516632INNOVA8, 530888ANG206, 541686SPR1, 561548VAS2, 561694IN1, 561776IN41, 561784INSP, 570321LAB2, 574294IN41, 574647IN2, 586573SP1, 6012884100SUR, 601376SP4100, 609757IN4100, 614722INNOVA1, 615284OR15, 615342CV7, 617636IN1, 618549CM4100, 620665INNOVA41, 623207SP, 623848SP, 630275G41, 630275G41B, 6309784100, 650493ANGIOCT, 662244GTSP, 678EJCIR1, 7023604100, 706651SP1, 706733SP1, 706774USP3, 708202VA41, 7087564100, 708783CL4100, 708783SP4100, 713375U4100, 7137764100, 718250ANGIO, 718630LIBERTY, 727398SPEC1, 731541RD, 740592INNOVA, 770736SP1, 787620IN4100, 8014794100, 802295VALCA, 803256SP, 804320CATH2B, 812238UN41, 812825INNOVA1, 813745INTV, 813972INTV, 815740SC4100, 817465AR4100, 817922BA4100, 843577CL2, 843674VAS1, 843674VAS2, 843789SP1, 843797TIR2, 8458584100, 847316INNOVA41, 850747IN4100, 858939IR9, 8593134100, 85931341002, 863687INV24, 864573LAB2, 865373SP, 865481SUR, 904BSSP, 906485ANGIO, 912819SP1, 913317SLS4100, 915577SP1, 915577SP2, 918494SFSP41, 918682IN41, 918744JR4100, 937352IN4100, 970247INNOVA, 989466INNOVA, COR359867, DR4100, MCL4100, PEMBROKE4100, RADNETXV1, and RADNETXV2.
510(k) numbers
["K052412"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register