Guidant Corporation: DISCOVERY II (model nos. 481, 981 1184, 1186, 1187, 1283, 1284, 1285 and 1286) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . DISCOVERY II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering both adaptive-rate therapy and conventional pacing therapy. These pacemakers provide various levels of therapeutic and diagnostic functionality. DISCOVERY II adaptive-rate models have an accelerometer, which is a motion sensor that responds to patient activity.
- Recall number
- Z-1036-05
- Recalling firm
- Guidant Corporation
- Product
- DISCOVERY II (model nos. 481, 981 1184, 1186, 1187, 1283, 1284, 1285 and 1286) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . DISCOVERY II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering both adaptive-rate therapy and conventional pacing therapy. These pacemakers provide various levels of therapeutic and diagnostic functionality. DISCOVERY II adaptive-rate models have an accelerometer, which is a motion sensor that responds to patient activity.
- Status
- Terminated
- Initiated
- 2005-07-18
- Posted
- 2005-07-28
- Terminated
- 2007-01-05
- Reason
- A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
- Root cause
- Other
- Action
- INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.
- Quantity
- INITIAL RECALL 414. EXPANDED RECALL 2,367
- Distribution
- worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Polynesia, Portugal, Reunion, Russia, Saudi Arabia, Singapore,Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan ROC, Turkey, United Kingdom, Uruguay and Venezuela. EXPANDED RECALL additional countries include: China, Croatia, Korea. Lithuania, Madagascar, Maldives, Malta, Monaco, Nepal and Thailand.
- Lot, serial or model codes
- Affected devices may be contained in the following serial number ranges. Model 481: INITIAL RECALL serial numbers 100000 thru 100115. EXPANDED RECALL serial numbers 100185, 100213, 100356, 100357, 100360. Model 981: INITIAL RECALL serial numbers 200002 thru 200041. EXPANDED RECALL serial numbers 200145, 200150, 200178, 200192, 200312. Model 1184: INITIAL RECALL serial numbers 300012 thru 300061. EXPANDED RECALL serial numbers 300159 thru 301072 and 610018 thru 610035. Model 1186: INITIAL RECALL serial numbers 500000 thru 500054. EXPANDED RECALL serial numbers 500056 thru 500444 Model 1187: INITIAL RECALL serial numbers 450000 thru 450010. EXPANDED RECALL serial numbers 450027 thru 450056. Model 1283: INITIAL RECALL serial numbers 600001 thru 600073. EXPANDED RECALL serial numbers 600214 thru 602143 Model 1284: INITIAL RECALL serial numbers 700000 thru 700079. EXPANDED RECALL serial numbers 700296, 700343, 700452, 700506, 700552, 700561, 700566. Model 1286: INITIAL RECALL serial numbers 900000 thru 900059. EXPANDED RECALL serial numbers 900045 thru 900888.
- 510(k) numbers
- not on file
- PMA numbers
- ["P940031"]
- Product code
- LWP
- Firm FEI
- 2124215
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28