Skip to the content

FDA Medical Device Recalls (openFDA)

Animas Corporation: Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Recall number
Z-1034-2015
Recalling firm
Animas Corporation
Product
Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.
Status
Terminated
Initiated
2011-09-06
Posted
2015-01-30
Terminated
2015-08-10
Reason
Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.
Root cause
Software design (manufacturing process)
Action
The field action was initiated on 8/29/2011 and comprised communication to patients and distributors using email, letters and verbal communication. In addition, notifications were sent to those Health Care Professionals who had patients that were affected by this field action. Health Authorities were notified in those countries where the pump had been distributed (France, Germany, Sweden and United Kingdom) or was in the hands of patients. Replacement product was provided to distributors and end users who were identified as having affected product.
Quantity
1235
Distribution
No US distribution, Distributors are located in France, Germany, Sweden and United Kingdom.
Lot, serial or model codes
Model Number(s):   100515-63 100510-63 100514-63 100512-63 100511-63 101200-03 100201-03 101202-03 101204-03 101205-03 101200-53 101202-53 101205-53 101200-02 101200-57 101201-57 101202-57 101204-57 101205-57 101200-63 101201-63 101202-63 101204-63 101205-63 101206-63
510(k) numbers
not on file
PMA numbers
["P130007"]
Product code
OYC
Firm FEI
3002920072
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register