Skip to the content

FDA Medical Device Recalls (openFDA)

ACell, Inc: Gentrix Surgical Matrix Thick, 20 x 30cm, Catalog Number PSMT2030, Lot 013948 - Product Usage: Gentrix Surgical Matrix Thick is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic and reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to the following procedures: hernia and body wall repair, colon and rectal prolapse repair, tissue repair, and esophageal repair.

Recall number
Z-1033-2019
Recalling firm
ACell, Inc
Product
Gentrix Surgical Matrix Thick, 20 x 30cm, Catalog Number PSMT2030, Lot 013948 - Product Usage: Gentrix Surgical Matrix Thick is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic and reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to the following procedures: hernia and body wall repair, colon and rectal prolapse repair, tissue repair, and esophageal repair.
Status
Terminated
Initiated
2019-02-14
Posted
not on file
Terminated
2020-05-07
Reason
The devices were released from a lot for which one sample failed to meet the company' s tensile strength specification for lot release.
Root cause
Nonconforming Material/Component
Action
Urgent Medical Device Recall notification letters dated 2/14/19 were sent to customers.
Quantity
1
Distribution
US nationwide distribution in the state of CA.
Lot, serial or model codes
SN KJ000423
510(k) numbers
["K170763"]
PMA numbers
not on file
Product code
FTM
Firm FEI
3005920706
Firm city
Columbia
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register