Skip to the content

FDA Medical Device Recalls (openFDA)

Meridian Bioscience Inc: PREMIER EHEC and PREMIER EHEC Bulk.

Recall number
Z-1033-2017
Recalling firm
Meridian Bioscience Inc
Product
PREMIER EHEC and PREMIER EHEC Bulk.
Status
Terminated
Initiated
2016-11-08
Posted
2017-01-17
Terminated
2017-08-16
Reason
Internal testing has demonstrated that the indicated kit lots produce false positive results when tested with some broth specimens.
Root cause
Under Investigation by firm
Action
On November 7, 2016 the firm sent Voluntary Medical Device Recall Letters and reply forms to their customers.
Quantity
4,747 EHEC + 180 EHEC Bulk
Distribution
Product was shipped to the following states: AL. AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WY. Product was also sent to the following countries: CANADA, GERMANY, HONG KONG, ITALY, JAPAN & NEW ZEALAND.
Lot, serial or model codes
PREMIER EHEC Product Code 608096: Lot # 608096H219, Exp Date 9/7/2017, Detection Antibody Reagent, Lot # 8603.138 Exp Date 9/30/2017; Lot # 608096H222, Exp Date 12/20/2017, Detection Antibody Reagent, Lot # 8603.14 Exp Date 1/12/2018; Lot # 608096H224, Exp Date 9/7/2017, Detection Antibody Reagent, Lot # 8603.141 Exp Date 1/29/2018; Lot # 608096H225, Exp Date 12/20/2017, Detection Antibody Reagent, Lot # 8603.142, Exp Date 2/5/2018 & Lot # 608096H226, Exp Date 3/6/2018, Detection Antibody Reagent, Lot # 8603.143, Exp Date 3/22/2018.  PREMIER EHEC Bulk Product Code 608096BU: Lot # 608096H220 Exp Date 9/7/2017, Detection Antibody Reagent, Lot # 8603.138, Exp Date 9/30/2017; Lot #, 608096H223, Exp Date 12/20/2017, Detection Antibody Reagent, Lot # 8603.14, Exp Date 1/12/2018 & Lot # 608096H227, Exp Date 3/6/2018, Detection Antibody Reagent, Lot # 8603.143, Exp Date 3/22/2018.
510(k) numbers
["K953362"]
PMA numbers
not on file
Product code
GNA
Firm FEI
1524213
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register