FDA Medical Device Recalls (openFDA)
Intelifuse Inc: InteliFUSE, Drill Bit, 1.50 mm, J-Latch, QTY: 1, REF: DB.150.JL, non-sterile, 1515 Poydras st., Ste 1490, New Orleans, LA 70112, MDCI Arundel House, 1 Liverpool Gardens, Worthing, West Sussex BN11 1SL, United Kingdom Intended Use: Creating of holes for implant placement.
- Recall number
- Z-1032-2009
- Recalling firm
- Intelifuse Inc
- Product
- InteliFUSE, Drill Bit, 1.50 mm, J-Latch, QTY: 1, REF: DB.150.JL, non-sterile, 1515 Poydras st., Ste 1490, New Orleans, LA 70112, MDCI Arundel House, 1 Liverpool Gardens, Worthing, West Sussex BN11 1SL, United Kingdom Intended Use: Creating of holes for implant placement.
- Status
- Terminated
- Initiated
- 2008-07-30
- Posted
- 2009-03-05
- Terminated
- 2011-06-22
- Reason
- Manufacturing specification for drill bit was incorrect which, when used in conjunction with the drill guide, could cause the production of metal shavings.
- Root cause
- Process control
- Action
- The firm initiated their recall on July 31, 2008 via telephone and Urgent Recall Notice to their current distributor, Integra. The letter asked them to complete, sign, and return the enclosed Recall Acknowledgment and Return Form. The letter also stated that InteliFUSE would travel to their designated Distribution Center(s) to ensure testing and replacement of all defective units once the form has been received. The firm sent follow-up Urgent Recall Notice letters on August 14, 2008 directing the distributor to recall from the hospitals to whom they distributed the product and to two hospitals which were shipped the products directly. The letter asked that use of the affected products be immediately ceased and that the any stock on hand be quarantined. The Recall Acknowledgment and Return Form should be completed, signed, and returned and the products should be returned to InteliFUSE for immediate replacement.
- Quantity
- 2133 units
- Distribution
- Nationwide Distribution - NY and NJ.
- Lot, serial or model codes
- Lot Nos.: 984266, 78993, N64730, 17262-01, 064798, and 82030.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HTW
- Firm FEI
- 3005509435
- Firm city
- New Orleans
- Firm state
- LA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28