Medtronic Neuromodulation: Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.

Recall number
Z-1031-2013
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.
Status
Terminated
Initiated
2013-02-27
Posted
2013-04-02
Terminated
2013-11-21
Reason
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi
Root cause
Nonconforming Material/Component
Action
Medtronic sent a "Urgent Medical Device Correction" letter dated January 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Medtronic has found through device testing that if the optional footswitch (Model 60883) is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equipment. The status messages on the display screen and associated audible alarms operate as intended, and would notify the clinician of any change in status as described in chapter 2 of the system manual. Enclosed you will find an insulating cap that should be placed over the footswitch connector port when the Model 8930 RF Generator is used without the footswitch. The footswitch connector port is located on the rear panel of the Model 8930 RF Generator as shown below. For further questions In the United States, contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 or your local product representative. Outside of the United States, contact your local product representative.
Quantity
848 (696 US, 152 OUS)
Distribution
Nationwide Distribution including Puerto Rico and all States in continental USA except ID, MT, and NH.
Lot, serial or model codes
all Model 8930 radio frequency (RF) generators are affected
510(k) numbers
["K113380"]
PMA numbers
not on file
Product code
GEI
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register