Skip to the content

FDA Medical Device Recalls (openFDA)

Aomori Olympus Co., Ltd.: NEEDLEMASTER 4MMx 21G UPPER, model no. NM-610L-0421 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.

Recall number
Z-1030-2021
Recalling firm
Aomori Olympus Co., Ltd.
Product
NEEDLEMASTER 4MMx 21G UPPER, model no. NM-610L-0421 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.
Status
Terminated
Initiated
2021-01-04
Posted
not on file
Terminated
2024-03-11
Reason
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Root cause
Under Investigation by firm
Action
On January 4, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the identified list of Endo-therapy devices may have a defective seal which may compromise the sterility of the product. Customers were asked to take the following actions: 1. Immediately assess any product you have in stock to identify endotherapy products with affected lot numbers listed in the recall letter. Cease use of product and quarantine any affected product. 2. Call your Olympus customer service representative at 1-888-524-7266 to obtain a Returned Goods Authorization so that you may return the product with no charge to you. Olympus will issue a credit or replacement to your facility for any returned product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0391 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this corrective action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter.
Quantity
7,400,000 (globally); 260,395 (US)
Distribution
US Nationwide distribution.
Lot, serial or model codes
03K, 04K, 08K, 05V, 06V, 07V, 08V, 09V, 0XV
510(k) numbers
["K153625"]
PMA numbers
not on file
Product code
FBK
Firm FEI
3003995201
Firm city
Kuroishi
Firm state
2 Chome
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register