Abbott Laboratories: AxSYM Myoglobin Reagent Pack; list 03E43; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064
- Recall number
- Z-1030-2007
- Recalling firm
- Abbott Laboratories
- Product
- AxSYM Myoglobin Reagent Pack; list 03E43; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064
- Status
- Terminated
- Initiated
- 2007-04-23
- Posted
- 2007-07-11
- Terminated
- 2008-05-18
- Reason
- Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Root cause
- Other
- Action
- Abbott initiated a field correction recall by letter dated 4/23/07 to all customers receiving the affected reagents. The accounts were informed of the failure of the AxSYM to open (actuate) some reagent packs with the affected lots. The accounts were requested to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading the reagent pack on the instrument; and discard any reagent packs that do not open properly. If the user observes a probe crash, prior to testing samples again, the user must perform a probe crash recovery following the procedure in the Operations Manual. If necessary, they need to replace the probe. If a probe crash recovery is not performed, subsequent results may be impacted. The customers were requested to retain a copy of the letter for their records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.
- Quantity
- 244 packs
- Distribution
- Worldwide, including USA, Puerto Rico, Argentina, Australia, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Hong Kong, Mexico, New Zealand, Singapore, South Korea, Taiwan, Thailand and Venezuela.
- Lot, serial or model codes
- list 03E43; lot 48718M100
- 510(k) numbers
- ["K983848"]
- PMA numbers
- not on file
- Product code
- DDR
- Firm FEI
- 1415939
- Firm city
- Abbott Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28