Smith & Nephew, Inc.: smith&nephew SMF(TM) STANDARD OFFSET STIKTITE(TM) COATED STEM 12/14 TAPER -1 SIZE, STEM, REF 713552510

Recall number
Z-1028-2019
Recalling firm
Smith & Nephew, Inc.
Product
smith&nephew SMF(TM) STANDARD OFFSET STIKTITE(TM) COATED STEM 12/14 TAPER -1 SIZE, STEM, REF 713552510
Status
Terminated
Initiated
2019-02-19
Posted
not on file
Terminated
2021-05-21
Reason
Complaints were received indicating that the outer label incorrectly identified the product as size "1" instead of "-1". The product contained within the package is size "-1"
Root cause
Packaging process control
Action
The firm notified its direct accounts by email and letter on 02/18/2019. The notice requested return of the recalled products. Distributors were directed to notify their customers of the recall and ensure that the actions are carried out.
Quantity
58 units
Distribution
US: CA, TN, NC, AR, IN, FL, MO, and, Canada, China, France, Germany, Great Britain, India, Italy, Japan, Mexico, Poland, South Korea, Spain
Lot, serial or model codes
Batch Numbers: 11AM15163U, 11AM15184T, 11CM01097U, 15MM06466, 16EM13018, 16EM13035 & 18CM08027
510(k) numbers
["K103256"]
PMA numbers
not on file
Product code
LPH
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register