Skip to the content

FDA Medical Device Recalls (openFDA)

Ascension Orthopedics, Inc: Ascension PIP instrument tray base, manufactured by Ascension Orthopedics, Austin, Texas

Recall number
Z-1025-04
Recalling firm
Ascension Orthopedics, Inc
Product
Ascension PIP instrument tray base, manufactured by Ascension Orthopedics, Austin, Texas
Status
Terminated
Initiated
2003-10-02
Posted
2004-07-20
Terminated
2004-06-17
Reason
Cracks in the tray due to excessive use can cause perforations in the outer wrap used to maintain instrument tray sterility.
Root cause
Other
Action
Recall was initiated via letter included with replacement trays begining in October 2003.
Quantity
107 trays
Distribution
Nationwide distrbution to IA, TX, OH, FL, IN, MN, CO, WA, CA, MO, and to a consignee in Ontario, Canada
Lot, serial or model codes
All trays remaining in the marketplace are subject to the recall.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEG
Firm FEI
3002788818
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register