Siemens Healthcare Diagnostics, Inc.: ADVIA Centaur Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10491445 All lots

Recall number
Z-1024-2022
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
ADVIA Centaur Enhanced Estradiol (eE2) 500T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10491445 All lots
Status
Open, Classified
Initiated
2022-03-03
Posted
not on file
Terminated
not on file
Reason
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Root cause
Under Investigation by firm
Action
Siemens issued an Urgent Medical Device Correction AIMC 22-03.A.US and Urgent Field Safety Notice AIMC 22-03.A.OUS on 3/3/22. Letter states reason, health risk and action to take: Customers must discontinue use of plasma specimens for testing with the Atellica IM and ADVIA Centaur eE2 assays until further notice. Customers may continue to use the Atellica IM eE2 and ADVIA Centaur eE2 assays with serum specimens. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Questions: contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative
Quantity
952 units
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Bulgaria, Burkina, Faso Country AG, Cyprus, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Greece, Guadeloupe, India, Ireland, Israel, Italy, Kazakhstan, Latvia, Lithuania, Macedonia, Malta, Morocco, Nepal, Netherlands, Norway, Palestinian Ter, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Vatican city.
Lot, serial or model codes
All lot numbers UDI: (01)00630414574646(10)14530079(17)20220304 (01)00630414574646(10)53553079(17)20220304 (01)00630414574646(10)54118081(17)20220521 (01)00630414574646(10)62860081(17)20220521 (01)00630414574646(10)70552083(17)20220805 (01)00630414574646(10)76807076(17)20220115 (01)00630414574646(10)93281076(17)20220115
510(k) numbers
not on file
PMA numbers
not on file
Product code
CHP
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register