Zyno Medical LLC: Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient.

Recall number
Z-1024-2013
Recalling firm
Zyno Medical LLC
Product
Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient.
Status
Terminated
Initiated
2011-06-13
Posted
2013-03-29
Terminated
2016-07-25
Reason
Update Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump
Root cause
Process design
Action
Zyno Medical initiated recall 6/13/11 to accounts advising of a Preventative Maintenance (PM) Schedule and a software upgrade Odometer to monitor counts of accumulative fluid volume infused by the unit. On 3/1/13, firm issued correction notice to users that have not had the software implemented on their unit. Units will be serviced at their scheduled PM. The notice identified the product, problem, and actions to be taken by the customers.
Quantity
2589 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
LOT # 20080416-zs5, 20080711-zs5, 20080826-zs5, 20081022-zs5, 20090126-zs5 20090320-zs5, 20090507-zs5, 20090612-zs5, 20090706-zs5,20090910-zs5 20091110-zs5, 20100106-C, 20100420-D, 20100729, 20101005-BJ 20100910-SH, 20101013-SH, 20101206-SH, 20110110-SH, 20110111-SH 20110204-SH, 20110210-SH, 20110308-SH, 20110505-SH
510(k) numbers
["K071545","K100705"]
PMA numbers
not on file
Product code
OPP
Firm FEI
3006575795
Firm city
Natick
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register