Medacta Usa Inc: Medacta International SA, M.U.S.T. Rod, REF 03.50.433, Size 5.5mm, Length 480mm, Type Straight, Mat Co-Cr-Mo The rods act as a connector between the different pedicle screws to create a stable construct for the stabilization and fusion of the lumbar and thoracic spine. Indications for Use: The M.U.S.T. pedicle screw system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5). The pedicle screws are connected by 5.5mm Titanium alloy or CoCrMo rods to create a stable construct. The construct is secured using a set screw made of CoCrMo.

Recall number
Z-1023-2017
Recalling firm
Medacta Usa Inc
Product
Medacta International SA, M.U.S.T. Rod, REF 03.50.433, Size 5.5mm, Length 480mm, Type Straight, Mat Co-Cr-Mo The rods act as a connector between the different pedicle screws to create a stable construct for the stabilization and fusion of the lumbar and thoracic spine. Indications for Use: The M.U.S.T. pedicle screw system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5). The pedicle screws are connected by 5.5mm Titanium alloy or CoCrMo rods to create a stable construct. The construct is secured using a set screw made of CoCrMo.
Status
Terminated
Initiated
2016-12-07
Posted
2017-01-13
Terminated
2017-11-15
Reason
Damaged packaging, lack of sterility assurance
Root cause
Packaging
Action
The firm, Medacta International, immediately notified all customers by phone on 12/07/16 and further notified via email on 12/9/16. The firm also issued an "URGENT- Field Safety Notice" dated 7/12/2016 to all its customers. The notice described the product, problem and actions to be taken. The customers were instructed to return all units to Medacta International SA and forward the notice to all people and organizations potentially affected. If you have any questions contact Regulatory Affairs Manager at +41 (0) 91 696 60 60.
Quantity
82 units
Distribution
Worldwide Distribution- US (nationwide) distribution to states of: TX and CO; and countries of: Australia, Belgium, Spain, German, Italy and Switzerland.
Lot, serial or model codes
Lot #s 125212,144216, 148146
510(k) numbers
["K121115"]
PMA numbers
not on file
Product code
KWP
Firm FEI
3006639916
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register