INNOKAS MEDICAL OY: VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx Only Innokas Medical Oy. Catalog numbers 2067980-002, 2067980-006, 2067980-010, and 2067980-014.

Recall number
Z-1022-2015
Recalling firm
INNOKAS MEDICAL OY
Product
VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx Only Innokas Medical Oy. Catalog numbers 2067980-002, 2067980-006, 2067980-010, and 2067980-014.
Status
Terminated
Initiated
2014-12-03
Posted
2015-01-28
Terminated
2015-04-24
Reason
If the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.
Root cause
Software design
Action
A notification letter was sent to the distributors on 12/3/2014. The letter requests the devices be returned to the manufacturer for repair.
Quantity
20 units
Distribution
Distributed in the US and Europe.
Lot, serial or model codes
SK514470022YA, SK514470023YA, SK514470024YA, SK514470020YA, SK514470021YA, SK514100003YA, SK514470019YA, SK514470001YA, SK514470017YA, SK514470018YA, SK514450021YA, SK514450022YA, SK514450019YA, SK514470004YA, SK514470005YA, SK514390001YA, SK514150002YA, SK514150004YA, SK514150005YA, SK514300002YA
510(k) numbers
["K133810"]
PMA numbers
not on file
Product code
MWI
Firm FEI
3007615597
Firm city
KEMPELE
Firm state
not on file
Firm country
Finland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register