Genetic Testing Institute,inc: Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).

Recall number
Z-1022-2013
Recalling firm
Genetic Testing Institute,inc
Product
Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).
Status
Terminated
Initiated
2013-02-04
Posted
2013-03-28
Terminated
2016-03-21
Reason
Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati
Root cause
Process control
Action
Hologic/Gen-Probe sent an Urgent Medical Device Recall letter dated February 4, 2013, via FedEx overnight deliver yto all affected customers. The letter identified affected product, stated the issue, and asked that product from the affected lot stop being used. Reported results for all samples tested using the affected lot should be reviewed. The referring physician should be notified that any negative patient result may be erroneous. Remaining inventory should be returned to Gen-Probe. A response form was also asked to be returned. For questions customers were instructed to contact Technical Support at 262-754-1000 or waukeshatechsupport@gen-probe.com. For questions regarding this recall call 262-754-1026.
Quantity
386 kits (370 US, 16 OUS)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.
Lot, serial or model codes
Kit Lot Number 3000389
510(k) numbers
["K053559"]
PMA numbers
not on file
Product code
LCO
Firm FEI
2183608
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register