FDA Medical Device Recalls (openFDA)
Genetic Testing Institute,inc: Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).
- Recall number
- Z-1022-2013
- Recalling firm
- Genetic Testing Institute,inc
- Product
- Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).
- Status
- Terminated
- Initiated
- 2013-02-04
- Posted
- 2013-03-28
- Terminated
- 2016-03-21
- Reason
- Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati
- Root cause
- Process control
- Action
- Hologic/Gen-Probe sent an Urgent Medical Device Recall letter dated February 4, 2013, via FedEx overnight deliver yto all affected customers. The letter identified affected product, stated the issue, and asked that product from the affected lot stop being used. Reported results for all samples tested using the affected lot should be reviewed. The referring physician should be notified that any negative patient result may be erroneous. Remaining inventory should be returned to Gen-Probe. A response form was also asked to be returned. For questions customers were instructed to contact Technical Support at 262-754-1000 or waukeshatechsupport@gen-probe.com. For questions regarding this recall call 262-754-1026.
- Quantity
- 386 kits (370 US, 16 OUS)
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.
- Lot, serial or model codes
- Kit Lot Number 3000389
- 510(k) numbers
- ["K053559"]
- PMA numbers
- not on file
- Product code
- LCO
- Firm FEI
- 2183608
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28