Zimmer Biomet, Inc.: VERSYS 10 INCH BEADED FC REV 15.0X250MM BWD LT & RT VERSYS BEADED FULLCOAT REVISION HIP PROSTHESIS

Recall number
Z-1021-2018
Recalling firm
Zimmer Biomet, Inc.
Product
VERSYS 10 INCH BEADED FC REV 15.0X250MM BWD LT & RT VERSYS BEADED FULLCOAT REVISION HIP PROSTHESIS
Status
Terminated
Initiated
2017-11-29
Posted
not on file
Terminated
2020-04-02
Reason
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Root cause
Packaging
Action
A similar recall was initiated in January 2016 to remove remaining inventory packaged in the old bag. Further evaluation identified additional lots packaged in the old bag that were not included in the January 2016 recall; therefore, resulting in this new recall for the additional lots packaged in the old bag. A firm, Zimmer Biomet, sent an "Urgent Medical Device Recall" notices dated November 29, 2017 and response forms to customers. The notice described the product, problem and actions to be taken. The Customers should take the following action: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. b. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com. c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additi
Quantity
13,227 in total
Distribution
Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
Lot, serial or model codes
Item Number: 00-7843-015-11 Lot Numbers: 62097294 11006832 62116495 61981821 Item Number: 00-7843-015-12 Lot Numbers: 62085294 11004177 62097244 61759259 62133149 61807097 77000821 61963730
510(k) numbers
["K030079"]
PMA numbers
not on file
Product code
LPH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register