FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: Reunion TSA Peg Alignment Sound Catalog #5901-1038 The Peg Alignment Sound (PIN 5901-1038, Revision A) is an instrument used in the ReUnion TSA shoulder system for confirming seating height and peg alignment of the TSA Self-Pressurizing Polyethylene Glenoid implant prior to final implantation into the glenoid vault. Its purpose is to provide assurance that the drilled holes for seating of the Glenoid implant have been properly prepared using the provided instrumentation
- Recall number
- Z-1020-2017
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- Reunion TSA Peg Alignment Sound Catalog #5901-1038 The Peg Alignment Sound (PIN 5901-1038, Revision A) is an instrument used in the ReUnion TSA shoulder system for confirming seating height and peg alignment of the TSA Self-Pressurizing Polyethylene Glenoid implant prior to final implantation into the glenoid vault. Its purpose is to provide assurance that the drilled holes for seating of the Glenoid implant have been properly prepared using the provided instrumentation
- Status
- Terminated
- Initiated
- 2016-12-02
- Posted
- not on file
- Terminated
- 2018-08-01
- Reason
- Reunion TSA Peg Alignment Sound broke during surgery.
- Root cause
- Device Design
- Action
- Stryker Orthopaedics sent an Urgent Medical Device Recall Notification letter dated December 2, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to inform users of the Urgent Medical Device Recall and forward the notice to all those individuals who need to be aware within their organization. Return all affected products available to: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07431 1360184/RA2016-112 Complete and sign the enclosed Business Reply Form and fax a copy to 855-261-3635 or email to SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com. Customers with questions were instructed to call 201-831-6693.
- Quantity
- 131 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Catalog #5901-1038
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28