Skip to the content

FDA Medical Device Recalls (openFDA)

Meridian Bioscience Inc: Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300

Recall number
Z-1019-2026
Recalling firm
Meridian Bioscience Inc
Product
Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300
Status
Open, Classified
Initiated
2025-12-05
Posted
2026-01-05
Terminated
not on file
Reason
The affected lots show a decline in performance over time, which may lead to false-negative results.
Root cause
Under Investigation by firm
Action
An "URGENT MEDICAL DEVICE CORRECTION" dated 12/1/25 was sent to customers. Required Actions: " Identify and segregate any remaining inventory of impacted product lots and suspend further use. " Destroy remaining inventory of impacted product lot in accordance with state, local, and/or federal regulations. o Contact Technical Service for replacement product. " Review previously obtained results from Revogene C. difficile listed Lots numbers. Any negative results obtained should be re-evaluated in conjunction with presented clinical findings to determine if retesting is required. " Complete and return the CONFIRMATION OF NOTIFICATION Form. Complete this form even if there is no remaining inventory. Actions to Be Taken By Meridian Bioscience, Inc.: Meridian Bioscience, Inc. will replace any unused inventory and provide support for any necessary retesting, including replacement test materials as needed. Contact Information: For any question regarding this notification please call Meridian Bioscience Technical Service at 1-800-343-3858, or email at FieldActionSupport@meridianbioscience.com.
Quantity
211 units
Distribution
US Nationwide distribution and the OUS country of Italy.
Lot, serial or model codes
UDI-DI: 00840733102349; Lot Numbers: 410300U195, 410300U196, 410300U197, 410300U201
510(k) numbers
["K172569"]
PMA numbers
not on file
Product code
OZN
Firm FEI
1524213
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register