Baxter Healthcare Renal Div: HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R, T5C8300, T5C8300R
- Recall number
- Z-1019-05
- Recalling firm
- Baxter Healthcare Renal Div
- Product
- HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R, T5C8300, T5C8300R
- Status
- Terminated
- Initiated
- 2005-06-21
- Posted
- 2005-07-19
- Terminated
- 2007-12-11
- Reason
- The screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an elcetric shock hazard.
- Root cause
- Other
- Action
- Baxter sent Urgent Device Correction letters worldwide to their customers on 6/21/05 and home patients on 6/22/05. The accounts were informed of teh risk of electrical shock from screws used to secure the Power Entry Module to the back of the HomeChoice system, and were instructed to always connect the power cord to the power entry module first, before plugging the power cord into a grounded electrical outlet. The users were advised to use caution to avoid contact the power entry module screws while powering on or powering off the device, and were informed that a new mounting bracket that providews more reliable grounding was installed in instruments processed through the Baxter service organization as of May 2005. All remaining instruments will receive the new mounting bracket by August 2006, and Baxter will contact the accounts to arrange for the upgrade. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 1.
- Quantity
- 20,133 units
- Distribution
- Nationwide, including Puerto Rico, and internationally to Argentina, Austria, Australia, Belgium, Canada, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Guatemala, Hong Kong, Ireland, India, Italy, Korea, Mexico, Malaysia, Nicaragua, the Netherlands, New Zealand, Peru, the Philippines, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, the United Kingdom, Uruguay and Venezuela.
- Lot, serial or model codes
- all serial numbers
- 510(k) numbers
- ["K012988"]
- PMA numbers
- not on file
- Product code
- FKX
- Firm FEI
- 3000210122
- Firm city
- Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28