Randox Laboratories, Limited: Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP2757
- Recall number
- Z-1018-2025
- Recalling firm
- Randox Laboratories, Limited
- Product
- Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP2757
- Status
- Open, Classified
- Initiated
- 2024-12-20
- Posted
- 2025-01-23
- Terminated
- not on file
- Reason
- Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
- Root cause
- No Marketing Application
- Action
- Randox Laboratories Ltd (manufacturer) extended recall to the distribution center in the USA via email on 12/20/24. The distributor will then contact the customer directly via email. Letter states reason for recall, health risk and action to take: Action to be taken: " Discontinue use of and discard any copies of Roche Cobas c501 ISA listing nmol/1 claims. " Ensure clinical results are no longer reported in nmol/1 using the Roche Cobas c501 nmol/1 claims. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. If you have any questions or concerns please contact Randox Technical Services.
- Quantity
- 33 kits
- Distribution
- US Nationwide distribution Including PR.
- Lot, serial or model codes
- UDI-DI: 05055273209198 All lots within expiry
- 510(k) numbers
- ["K011568"]
- PMA numbers
- not on file
- Product code
- DFC
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28