Skip to the content

FDA Medical Device Recalls (openFDA)

St Jude Medical: Implantable Cardioverter Defibrilators Epic DR/HF (V-233/V-337/V-338), Epic Plus DR/VR/HF (V-236/V-239/V-196/V-239T/V-196T/V-350) Atlas DR (V242), and Atlas Plus DR/VR/HF (V-243/V-193C/V-340/V-341/V-343).

Recall number
Z-1017-05
Recalling firm
St Jude Medical
Product
Implantable Cardioverter Defibrilators Epic DR/HF (V-233/V-337/V-338), Epic Plus DR/VR/HF (V-236/V-239/V-196/V-239T/V-196T/V-350) Atlas DR (V242), and Atlas Plus DR/VR/HF (V-243/V-193C/V-340/V-341/V-343).
Status
Terminated
Initiated
2005-06-17
Posted
2005-07-19
Terminated
2012-03-26
Reason
Two corrections to firmware due to Skipped-Charge Shock and Sensor Noise Anomaly are being made with one noninvasive upgrade.
Root cause
Other
Action
Recall was by letter to Physicians sent by FEDEX. Follow-up upgrade to programmers will be done by firm representatives to the new version 4.8.5.
Quantity
30,000 domestically, 11,684 internationally
Distribution
Nationwide and Sweden, Canada, Australia, New Zealand. South America and Asia
Lot, serial or model codes
No serial numbers were given. All models mentioned are affected.
510(k) numbers
not on file
PMA numbers
["P910023/S77","P030054/S7"]
Product code
NIK
Firm FEI
2017865
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28