St Jude Medical: Implantable Cardioverter Defibrilators Epic DR/HF (V-233/V-337/V-338), Epic Plus DR/VR/HF (V-236/V-239/V-196/V-239T/V-196T/V-350) Atlas DR (V242), and Atlas Plus DR/VR/HF (V-243/V-193C/V-340/V-341/V-343).
- Recall number
- Z-1017-05
- Recalling firm
- St Jude Medical
- Product
- Implantable Cardioverter Defibrilators Epic DR/HF (V-233/V-337/V-338), Epic Plus DR/VR/HF (V-236/V-239/V-196/V-239T/V-196T/V-350) Atlas DR (V242), and Atlas Plus DR/VR/HF (V-243/V-193C/V-340/V-341/V-343).
- Status
- Terminated
- Initiated
- 2005-06-17
- Posted
- 2005-07-19
- Terminated
- 2012-03-26
- Reason
- Two corrections to firmware due to Skipped-Charge Shock and Sensor Noise Anomaly are being made with one noninvasive upgrade.
- Root cause
- Other
- Action
- Recall was by letter to Physicians sent by FEDEX. Follow-up upgrade to programmers will be done by firm representatives to the new version 4.8.5.
- Quantity
- 30,000 domestically, 11,684 internationally
- Distribution
- Nationwide and Sweden, Canada, Australia, New Zealand. South America and Asia
- Lot, serial or model codes
- No serial numbers were given. All models mentioned are affected.
- 510(k) numbers
- not on file
- PMA numbers
- ["P910023/S77","P030054/S7"]
- Product code
- NIK
- Firm FEI
- 2017865
- Firm city
- Sylmar
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28