Skip to the content

FDA Medical Device Recalls (openFDA)

Future Diagnostics Solutions B.V.: Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.

Recall number
Z-1016-2023
Recalling firm
Future Diagnostics Solutions B.V.
Product
Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.
Status
Open, Classified
Initiated
2022-12-06
Posted
2023-01-25
Terminated
not on file
Reason
Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.
Root cause
Under Investigation by firm
Action
On 12/06/2022, the firm send a Recall Notice via email to customer informing them that The PTH kit 4K-IPT-00 with lot numbers: M78559, M78560 and M78636 can occasionally give a high duplicate %CV outside the recommended maximum of 15%CV. As a result of the failure in the duplicate CV requires a repeat run it may cause a delay in the patient result turnaround time or in system calibration depending on what samples are run and can lead to a wrong sample result. Actions to be taken by customers: -Immediately identify and quarantine the affected lot numbers. -Further instructions will be provided by the Future Diagnostics. For questions, send an email to fieldsafety@future-diagnostics.nl.
Quantity
342 kits
Distribution
Worldwide distribution. US states of AZ, CA, CO, LA, OH, PA, and TX; Austria, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Spain, Sweden, and United Kingdom
Lot, serial or model codes
Catalog Number: 4K-IPT-00 UDI-DI: 08719326045214-230915 Lot Numbers: M78559 M78560 M78636
510(k) numbers
["K000471"]
PMA numbers
not on file
Product code
CEW
Firm FEI
3001857593
Firm city
Wijchen
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register