Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc.: Minibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device.

Recall number
Z-1016-2017
Recalling firm
Hospira Inc.
Product
Minibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device.
Status
Terminated
Initiated
2013-02-19
Posted
2017-01-11
Terminated
2017-04-24
Reason
Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.
Root cause
Process control
Action
Hospira sent an Urgent Device Recall letter dated February 19, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please check your inventory and immediately quarantine any affected product. Complete the attached reply form and return it to the fax number or e-mail address on the form, even if you do not have the affected product. Inform healthcare professionals in your organization of this recall. Return affected product to Stericycle using the label provided with this letter. If you have not received a return label or require additional assistance contact Stericycle at 1-888-965-5821 between the hours of Bam to 5pm EST, Monday through Friday. To ensure proper and timely credit, follow the instructions on the return label for returning the product. Please visit http://expertezlabel.com to request additional labels for returning affected product. If you have distributed the product further, notify your accounts that received the product identified above of this recall and ask them to contact Stericycle to receive a reply form and return labels for returning the product. Please contact Hospira Customer Care at 1-877-946-7747, between the hours of 6am to 6pm CST, Monday through Friday, or your Hospira representative regarding replacement product availability. For further questions please call (224) 212-2000.
Quantity
48,050 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Product List Number: 14699-28;  Lot Number: 17-077-NS
510(k) numbers
["K941214"]
PMA numbers
not on file
Product code
FPK
Firm FEI
3013319212
Firm city
Lake Forest
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register