Mobius Imaging, LLC: Airo Mobile Intraoperative CT I Airo; Model #: MobiCT-32. Computed Tomography x-ray system.

Recall number
Z-1016-2015
Recalling firm
Mobius Imaging, LLC
Product
Airo Mobile Intraoperative CT I Airo; Model #: MobiCT-32. Computed Tomography x-ray system.
Status
Terminated
Initiated
2014-12-08
Posted
2015-01-27
Terminated
2015-07-06
Reason
There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect navigation data being transferred to the curve.
Root cause
Software design
Action
Mobius sent an Urgent Field Safety Notice dated December 8, 2014, to all affected customers. The letter explains the recall and provides advise on action to be taken by the user. The letter also informs consignees that Mobius Imaging is working on a software fix for the software bug/communication error. The final fix is anticipated in early 2015. Advise on action to be taken by the user: The user should be prepared for the potential necessity to perform a manual navigation registration in order to use navigation. In general, do not use navigation data transmitted by the AIRO Mobile CT Scanner of any scan the Brainlab AIR App displayed the "Received Wrong Dataset" error for. To use such scans with the Curve, a manual navigation registration is required. To avoid the possibility of incorrect data from being transferred to the Curve, Mobius Imaging is recommending that customers only select "NO" when they come to the "Continue with same patient" screen after each scan. They can select the same patient under the "Existing Patients" tab and restart the workflow. Additionally, always adequately verify the accuracy of the navigation registration data as required by the Brainlab Curve" Navigation system software and the AERO Operator's Manual. If you require further clarification, please feel free to contact your local Brain lab Customer Support Representative. Customer Hotline: +49 89 99 15 68 44 or +1800 597 5911 (for US customers) or by E-mail: support@brainlab.com (for US customers: us.support@brainlab.com) Fax Brainlab AG: + 49 89 99 15 68 33 Address: Brainlab AG (headquarters), Kapellenstrasse 12, 85622 Feldkirchen, Germany. For questions regarding this recall call 978-615-5025.
Quantity
21 devices distributed. 15 of these devices affected by the software v. 1.1.1+ patch
Distribution
USA including AK, IL, WA, MA, NY, WV, NC, TX, CA, PA and Internationally to Germany, Netherlands, Belgium, Switzerland, Saudi Arabia, Egypt.
Lot, serial or model codes
Serial #s: 2251 (#1)*, 2324 (#2), 2328(#3), 2344 (#4), 2346 (#5), AIRO-0106, AIRO-0107*, AIRO-0108, AIRO-0110, AIRO-0111, AIRO-0112, AIRO-0113, AIRO-0114, AIRO-0115, AIRO-0116, AIRO-0117, AIRO-0118, AIRO-0119, AIRO-0120, AIRO-0122, and AIRO-0123. *Demo system
510(k) numbers
["K131431"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3010151377
Firm city
Ayer
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register