Skip to the content

FDA Medical Device Recalls (openFDA)

Biogenex Laboratories: Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System; Model Number/Catalog Number: AS4010B; The product is manufactured by: BioGenex Laboratories, Inc. 4600 Norris Canyon Road San Ramon, CA 94583 and Distributed by: Abbott Molecular, 1300 E. Touhy Ave - 2E, Des Plaines, IL 60018 Xmatrx Automated Staining System is intended for in vitro diagnostic use. It is suitable for fluorescent hybridization (FISH) applications.

Recall number
Z-1016-2010
Recalling firm
Biogenex Laboratories
Product
Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System; Model Number/Catalog Number: AS4010B; The product is manufactured by: BioGenex Laboratories, Inc. 4600 Norris Canyon Road San Ramon, CA 94583 and Distributed by: Abbott Molecular, 1300 E. Touhy Ave - 2E, Des Plaines, IL 60018 Xmatrx Automated Staining System is intended for in vitro diagnostic use. It is suitable for fluorescent hybridization (FISH) applications.
Status
Terminated
Initiated
2009-01-23
Posted
2010-03-08
Terminated
2010-12-02
Reason
Under certain conditions, product system failure has the potential to cause operator injury due to splash of Hemo-De solution.
Root cause
Device Design
Action
On 1/23/09, the firm issued an Advisory Notice to all its direct consignees, informing them of the affected product and providing instructions on the recall. Please contact BioGenex Technical Support at 1-800-421-4149 for any assistance you may need.
Quantity
12 units
Distribution
The firm distributed the product to one (1) distributor as a consignee of our product. The firm indicated that no customer has received the product directly from BioGenex, due to the fact that the product is exclusive distributed by Abbott Molecular.
Lot, serial or model codes
Serial Numbers: AS40211, AS40226, AS40230, AS40231, AS40232, AS40233, AS40234, AS40235, AS40236, and AS40237
510(k) numbers
not on file
PMA numbers
not on file
Product code
KEY
Firm FEI
2936532
Firm city
San Ramon
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register