Siemens Healthcare Diagnostics, Inc.: IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]

Recall number
Z-1015-2022
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]
Status
Open, Classified
Initiated
2022-02-22
Posted
not on file
Terminated
not on file
Reason
Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.
Root cause
Under Investigation by firm
Action
Urgent Medical Device Correction was issued February 02, 2022 via Fed Ex. Notification provides instructions to customers to mitigate this sample carryover. Customers are requested to complete and return the Effectiveness Check questionnaire. Contact your local Siemens Healthineers technical support representative with any questions.
Quantity
10,755 boxes (533 US, 10,222 OUS)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, ID, IN, MN, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TX, VA, WI, WY and OUS: 81 countries.
Lot, serial or model codes
8 production lots: 463, 464, 466, 467, 469, 470, 471, 472, [UDI for 200 tests: (01)00630414961132(10)463(17)20220228, (01)00630414961132(10)464(17)20220331, (01)00630414961132(10)466(17)20220331, (01)00630414961132(10)467(17)20220531, (01)00630414961132(10)469(17)20220731, (01)00630414961132(10)470(17)20220930, (01)00630414961132(10)471(17)20221130, (01)00630414961132(10)472(17)20230228] [UDI for 600 tests: (01)00630414961149(10)463(17)20220228, 01)00630414961149(10)464(17)20220331, (01)00630414961149(10)466(17)20220331, (01)00630414961149(10)467(17)20220531, (01)00630414961149(10)469(17)20220731, (01)00630414961149(10)470(17)20220930, (01)00630414961149(10)471(17)20221130, (01)00630414961149(10)472(17)20230228]
510(k) numbers
["K990222"]
PMA numbers
not on file
Product code
DHA
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register