Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: Infant Warmer System (IWS)

Recall number
Z-1015-2017
Recalling firm
GE Healthcare, LLC
Product
Infant Warmer System (IWS)
Status
Terminated
Initiated
2016-08-04
Posted
not on file
Terminated
2018-07-12
Reason
GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
Root cause
Use error
Action
This recall strategy includes the depth of the recall, plan for public warning and effectiveness check. The strategy outlined below addresses these elements for this particular recall. The plan for public warning is to contact all customers with potentially affected product by means of an Urgent Medical Device Correction letter. The letter will instruct customers that they can continue to use the device by following instructions on the Addendum to the Service Manual. The Service Manual Addendum provides instructions for checking and tightening the screws. In the US the letter will be sent by a traceable means [either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX]. This letter will be sent to: Nurse Manager, Labor & Delivery/NICU; Manager, Bio-Medical Engineering; and Director of Risk Management.
Quantity
32,073 (US = 24,361; OUS = 7,712)
Distribution
This product has been distributed domestically and internationally .
Lot, serial or model codes
Medical device listing number: D103053, D103054, D103056, D103057, D183529
510(k) numbers
["K963058"]
PMA numbers
not on file
Product code
FMT
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register