FDA Medical Device Recalls (openFDA)

Warsaw Orthopedic, Inc. dba Medtronic SOFAMOR DANEK: Medtronic TSRH Spinal System, 3Dx Splined Connector, Small; Medtronic Sofamor Danek, Memphis, TN; REF 8351510. The device is intended for orthosis, spinal pedicle fixation for degenerative disc disease. The product is used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients; posterior non-cervical, non-pedicle screw fixation or for anterior use for certain conditions.

Recall number
Z-1015-2009
Recalling firm
Warsaw Orthopedic, Inc. dba Medtronic SOFAMOR DANEK
Product
Medtronic TSRH Spinal System, 3Dx Splined Connector, Small; Medtronic Sofamor Danek, Memphis, TN; REF 8351510. The device is intended for orthosis, spinal pedicle fixation for degenerative disc disease. The product is used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients; posterior non-cervical, non-pedicle screw fixation or for anterior use for certain conditions.
Status
Terminated
Initiated
2008-10-31
Posted
2009-03-12
Terminated
2009-11-18
Reason
The thickness of the connector may be under specification, which could cause the implant to break during surgery or after implantation.
Root cause
Process control
Action
Hospital risk managers and implanting physicians were notified separately by letters dated 1/9/09. The letters advised of the recall, reported that a firm representative would visit the facility to pick up the recalled products. Physicians were instructed to closely monitor their patients during postoperative visits, to notify any other physician who may be monitoring the patient in lieu of the implanting physician, and to advise their patients to contact them immediately if they have any sudden and significant symptoms (increased discomfort, sudden change in their clinical symptoms, or new symptoms such as leg pain, back pain or irritation of local tissue). Customers were also instructed to complete and return the attached questionnaire to Medtronic SOFAMOR DANEK via mail (enclosed self-addressed stamped envelope) or fax at 901-399-2047. Direct questions to your Medtronic SOFAMOR DANEK Sales Representative or the firm's Global Quality Department at 800-876-3133, extension 6333.
Quantity
3,545
Distribution
Nationwide Distribution including states of Arizona, Connecticut, Georgia, Indiana, Minnesota, New Jersey, New York, South Carolina, Texas and Washington.
Lot, serial or model codes
All lots.
510(k) numbers
["K072317"]
PMA numbers
not on file
Product code
MNI
Firm FEI
1824199
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register