FDA Medical Device Recalls (openFDA)
Cardiopulmonary Instrumentation, Inc: CI healthcare***Model: CI8092***S/N: 5-60-85-044 MFD: JUN, 2010 MADE IN CHINA***cihealthcare.com This stainless steel stretcher is used for transporting patients.
- Recall number
- Z-1013-2012
- Recalling firm
- Cardiopulmonary Instrumentation, Inc
- Product
- CI healthcare***Model: CI8092***S/N: 5-60-85-044 MFD: JUN, 2010 MADE IN CHINA***cihealthcare.com This stainless steel stretcher is used for transporting patients.
- Status
- Terminated
- Initiated
- 2011-02-17
- Posted
- 2012-02-14
- Terminated
- 2012-04-19
- Reason
- Cardiopulmonary Instrumentation, Inc. dba CI Healthcare, Miami, FL, is recalling their Stainless Steel Patient Transport Stretcher, model # CI-8092, Lot 002, because of a malfunction that occurs when weight is applied to the backrest. This malfunction causes the backrest support bar, when secured in the first hook, to fall to a flat position.
- Root cause
- Device Design
- Action
- CI Healthcare sent an "URGENT RECALL NOTIFICATION" letter dated February 17, 2011, to its customer. The letter describes the product, problem and actions to be taken. The customer were instructed to do the following: 1) Examine their inventory and isolate the product(s). 2) If they have distributed any of the products to other services or facilities, please forward this information as appropriate. 3) Discontinue use of the product until a correction has been properly executed and 4) Complete and return the "Recall Response Form" as soon as possible. Should you have any questions, please call Robert Nunez at (305) 592-8196.
- Quantity
- 72 Units
- Distribution
- Nationwide Distribution - to one distributor in Puerto Rico
- Lot, serial or model codes
- Catalog # CI-8092, Lot # 002, Serial #'s: 5-02-92-489, 5-60-85-044, 5-02-40-687, 5-39-51-261, 5-87-72-685, 5-70-66-682, 5-83-08-847, 5-68-87-204, 5-23-81-803, 5-09-48-748, 5-46-83-142, 5-97-10-627, 5-40-10-097, 5-58-95-742, 5-42-38-793, 5-01-48-417, 5-57-51-677, 5-31-07-403, 5-41-68-771, 5-55-89-200, 5-02-04-936, 5-00-36-598, 5-29-04-097, 5-33-44-600, 5-44-36-010, 5-50-54-239, 5-08-76-966, 5-45-28-563, 5-28-92-924, 5-70-91-773, 5-70-20-478, 5-47-73-702, 5-14-47-739, 5-06-38-206, 5-60-94-228, 5-44-44-162, 5-86-06-465, 5-76-88-888, 5-72-01-700, 5-88-02-542, 5-83-56-148, 5-03-63-494, 5-84-23-417, 5-87-09-630, 5-32-40-372, 5-49-44-489, 5-01-76-666, 5-40-08-584, 5-91-42-591, 5-32-95-591, 5-49-12-189, 5-29-86-200, 5-58-40-808, 5-84-08-385, 5-04-69-827, 5-71-18-330, 5-13-18-375, 5-74-67-486, 5-08-13-310, 5-04-26-945, 5-76-35-146, 5-43-18-883, 5-61-74-389, 5-02-88-727, 5-12-31-931, 5-10-62-445, 5-90-40-312, 5-83-76-170, 5-64-45-267, 5-54-92-118, 5-53-46-840, 5-77-99-209.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FPO
- Firm FEI
- 1034275
- Firm city
- Doral
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28