FDA Medical Device Recalls (openFDA)
Bio-Rad Laboratories Inc: Bio-Rad VARIANT II Hemoglobin A1c Program Recorder Pack with Calibrator/Diluent Set, Model Number 270-2101NU Manufactured by Bio-Rad Laboratories, Hercules, CA
- Recall number
- Z-1013-2010
- Recalling firm
- Bio-Rad Laboratories Inc
- Product
- Bio-Rad VARIANT II Hemoglobin A1c Program Recorder Pack with Calibrator/Diluent Set, Model Number 270-2101NU Manufactured by Bio-Rad Laboratories, Hercules, CA
- Status
- Terminated
- Initiated
- 2009-06-15
- Posted
- 2010-03-05
- Terminated
- 2011-02-02
- Reason
- Calibration Failure-- due to high intercept values, the device fails calibration and users are unable to run samples.
- Root cause
- Other
- Action
- Recall initiated June 15, 2009, with notification initially conducted by phone with Fax Notification follow up domestically. Each subsidiary will be given an urgent notice, subsidiary response form, customer notification, customer acknowledgement form. Each subsidiary is to contact customers and translate notification and response form as needed. Contact the recalling firm for additional information.
- Quantity
- 613 units
- Distribution
- Product was distributed to 127 customers throughout the US and internationally to Argentina, Australia, Canada, Chile, Czech Republic, France, Germany, Hong Kong, Hungary, India, Italy, New Zealand, Poland, Singapore, Sweden, Taiwan, and the United Kingdom
- Lot, serial or model codes
- Medical device, intended for percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography, intended for professional use only.
- 510(k) numbers
- ["K070452"]
- PMA numbers
- not on file
- Product code
- LCP
- Firm FEI
- 2915274
- Firm city
- Hercules
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28