DePuy Orthopaedics, Inc.: The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Product labeling: LEO-1100-30-000 REV D.pdf.pdf, LEO-1100-40-000 REV D.pdf.pdf, LEO-110030100_REV D.pdf.pdf, LEO-110030110_REV D.pdf.pdf, LEO-110030120_REV D.pdf.pdf

Recall number
Z-1012-2018
Recalling firm
DePuy Orthopaedics, Inc.
Product
The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Product labeling: LEO-1100-30-000 REV D.pdf.pdf, LEO-1100-40-000 REV D.pdf.pdf, LEO-110030100_REV D.pdf.pdf, LEO-110030110_REV D.pdf.pdf, LEO-110030120_REV D.pdf.pdf
Status
Terminated
Initiated
2017-11-28
Posted
2018-02-27
Terminated
2020-03-27
Reason
The screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays.
Root cause
Employee error
Action
DePuy Orthopaedics, Inc. is voluntarily recalling 11 lots of the GLOBAL UNITE Platform Shoulder System. The device is being recalled because the screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays. Once the screw is properly assembled, there is no additional risk to the patient. Further distribution or use of the affected lots is to cease immediately.DePuy Orthopaedics, Inc. is not recommending prophylactic revision. The company recommends that surgeons discuss potential clinical implications and risks with symptomatic patients that received the affected lots and may have experienced a surgical delay. Please take the following actions: Please cease using the affected components immediately. Return affected devices: Medical facilities are to determine if any of the recalled devices are still on hand, and return affected devices immediately to their DePuy Synthes Sales Consultant or return them to DePuy Synthes Joint Reconstruction for credit following normal procedures. Reconciliation forms should be faxed to 574-371-4939. Note: The affected lots may be on consignment at your facility. Reconciliation Form: Complete the Reconciliation Form and return to your DePuy Synthes Sales Consultant, email to DPYUS-JointReconFieldActions@its.jnj.com, or fax to 574-371-4939 within five (5) days of this notice. Records: Retain a copy of the completed Reconciliation Form in your files along with this notice. Additional Notifications: Notify surgeons at your facility by providing them with a copy of this notice to ensure surgeons are aware of this recall. Forward this notice to others in your facility that need to be informed.If any affected product has been forwarded to another facility, contact that facility immediately to communicate this field action with the facility/facilities. Maintain a copy of this notice wi
Quantity
85
Distribution
Distribution Level: The device removal is extended to the companies US Distributors/Sales Offices and Medical Facilities that may have used, purchased, or received the affected devices. Approximately 53 units distributed to firm's international affiliates.
Lot, serial or model codes
Lot # 8556802, 8556803, 8556804, 8556805, 8583381, 8556910, 8556951, 8605849, 8564487, 8572431, 8610357, 8527674.
510(k) numbers
["K133834"]
PMA numbers
not on file
Product code
KWS
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register