Skip to the content

FDA Medical Device Recalls (openFDA)

Aesculap, Inc.: Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Recall number
Z-1012-2015
Recalling firm
Aesculap, Inc.
Product
Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.
Status
Terminated
Initiated
2014-12-17
Posted
2015-01-26
Terminated
2016-12-13
Reason
Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).
Root cause
Labeling False and Misleading
Action
Aesculap sent an Important Correction and Removal and Field Safety Notification letter, dated December 17, 2014, to all affected customers. The letter identiifed the affected product, problem, and actions to be taken. Customers were requested to remove from inventory and contact Aesculap regarding return of the affected devices and acknowledgement form.
Quantity
772
Distribution
Worldwide Distribution - US Nationwide
Lot, serial or model codes
material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T
510(k) numbers
["K120559"]
PMA numbers
not on file
Product code
JXG
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register