FDA Medical Device Recalls (openFDA)
Angiodynamics, Inc.: PulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20, UPN: H787125004045, REF: 12500404; Pulse Spray 3Fx135x20, UPN: H787125004155, REF: 12500415; and PULSE SPRAY 5F/45/10, UPN: H787125004235, REF: 12500423.
- Recall number
- Z-1010-2024
- Recalling firm
- Angiodynamics, Inc.
- Product
- PulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20, UPN: H787125004045, REF: 12500404; Pulse Spray 3Fx135x20, UPN: H787125004155, REF: 12500415; and PULSE SPRAY 5F/45/10, UPN: H787125004235, REF: 12500423.
- Status
- Open, Classified
- Initiated
- 2023-12-18
- Posted
- 2024-02-07
- Terminated
- not on file
- Reason
- Rotating adaptor O-ring manufacturing defect impacting seal integrity which may lead to a leak during infusion of medication.
- Root cause
- Process control
- Action
- An Urgent Medical Voluntary Medical Device notices dated December 18, 2023 was issued to distributors and end users with instructions to immediately stop using affected product, remove said product from inventory and return to AngioDynamics, Inc. The notice should be forwarded to all site affected product may have been distributed to and recipients are asked to complete and return the Reply Verification Tracking Form. If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday - Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product.
- Quantity
- 7 eaches
- Distribution
- United States (MD, LA & UT), Austria, Canada and Spain.
- Lot, serial or model codes
- UDI-DI: 15051684013109, REF: 12500404, Lot: 5795345; UDI-DI: 15051684013215, REF: 12500415, Lot: 5793480; UDI-DI: 15051684013291, REF: 12500423, Lot: 5795344.
- 510(k) numbers
- ["K163356"]
- PMA numbers
- not on file
- Product code
- KRA
- Firm FEI
- 1319211
- Firm city
- Queensbury
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28